Disconnect the hose leading to evaporative purge valve from the charcoal
canister at the point of connection to charcoal canister to avoid entry of gas fumes into the engine. The hose must be blocked
to avoid entry or air as well. Reconnect after the procedure.
Warm up the engine and then stop it. Connect the jumper wires with the
voltmeter and the switch as shown. The switch must be in disconnected state.
Turn the screw of the AFM in all the way till it stops. Count 3 turns
as you turn now counter-clockwise. This is the base setting of the screw.
Restart the car. Observe the voltmeter display. There are three scenarios
one of which can take place:
1 - Various readings that switch fast and haphazardly between about .8V
and about .2V
2 - A relatively steady reading above .5V
3 - A relatively steady reading below .5V
Solutions:
1 - Turn the AFM screw one full revolution clockwise richening
the mixture. You should see the reading turn steady above .5V. If the reading does not turn steady above .5V the mixture is
too lean - turn the cogwheel of the AFM counter-clockwise, by one cog. Reset the screw to base setting - three full turns
out from full stop, and repeat the test.
When the reading does set steady above .5V the "rich" margin is
good.
Now turn the screw back to base setting, and then turn it once again
- one revolution counter-clockwise. You should see the reading turn relatively steady below .5V. If the reading does not turn
steady the mixture is too rich - turn the cogwheel of the AFM clockwise by one cog. Reset the screw to base setting -
three full turns out from stop, and repeat the test.
When the AFM is adjusted for crude operation within the band of
operation of the oxygen sensor, turning the screw clockwise one turn from base setting provides relatively
steady reading above .5V, turning the screw counter-clokwise one turn from base setting provides relatively steady reading
below .5V. When this condition is satisfied the AFM is adjusted - the screw must be set to base setting before adjustment
with gas analyzer.
At base setting the voltage fluctuates haphazardly as a result of engine
control by mapped data - the system is not able to keep the mixture at steady state with disconnected oxygen sensor due
to the fact that any given stroke is not same as the next one in miscroscopic view. That's the reason why oxygen sensor is
used.
2 - A relatively steady reading above .5V signifies rich condition as
mentioned just above. Turn the cogwheel of the AFM clockwise by one cog, to allow more air at given position to make the mixture
leaner. Reset the screw to base setting - three turns out from full stop and repeat the test.
3 - A relatively steady reading below .5V signifies lean
condition as mentioned just above. Turn the cogwheel of the AFM counter-clockwise by one cog, to allow less air at given position
to make the mixture richer. Reset the screw to base setting - three turns out from full stop and repeat the test.